화학공학소재연구정보센터
Applied Surface Science, Vol.258, No.14, 5399-5404, 2012
TiO2 nanotube arrays deposited on Ti substrate by anodic oxidation and their potential as a long-term drug delivery system for antimicrobial agents
Nanotube arrays on medical titanium surfaces were fabricated by two different anodization methods and their potential for storage and release of antimicrobial substances was evaluated. The treatment of the Ti surfaces in fluoride containing electrolytes on water as well as on polyethylene glycol basis led to the formation of TiO2 nanotubes with up to 6.54 mu m length and average diameters of up to 160 nm. Drug release experiments with the model antibiotic vancomycin and with antibacterial silver ions showed that the increased surface area of the anodized samples enabled them to be loaded with up to 450% more active agent than the untreated Ti surfaces. Significant surface-dependent differences in the release kinetics of vancomycin were observed. In comparison to surfaces anodized in an aqueous electrolyte, the release of the antibiotic from surfaces anodized in an electrolyte based on ethylene glycol was significantly retarded, with a release of noticeable amounts over a period of more than 300 days. Loading of nanotube surfaces fabricated in aqueous electrolyte with silver ions revealed increased amounts of adsorbed silver by up to 230%, while the release kinetics showed significant differences in comparison to untreated Ti. It was concluded that nanotube arrays on favored medical implant materials have a high potential for loading with antimicrobial agents and also provide the possibility of tailored release kinetics by variation of anodization parameters. (C) 2012 Elsevier B. V. All rights reserved.