Thermochimica Acta, Vol.522, No.1-2, 72-76, 2011
Anomalous transitions of DODAB using fast scanning liquid calorimetry
Anomalous melting transitions are observed on dimethyl dioctadecyl ammonium bromide (DODAB) using a fast-scanning liquid calorimeter. A single transition is observed at low scan rates (1 K/min) while up to four peaks are evident at high scan rates (60 K/min). These deconvoluted transitions are signatures of the basic properties of the transition kinetic landscape which are convoluted into a single peak at low scanning rates. The technique is also applied to the proteins Rnase A, Lysozyme and Cytochrome P450. The calorimeter is a glass capillary uniformly coated with Ag, which acts both as the heating element and resistive thermometer. (C) 2011 Elsevier B.V. All rights reserved.