화학공학소재연구정보센터
Langmuir, Vol.27, No.11, 6712-6719, 2011
Reliable Measurements of Interfacial Slip by Colloid Probe Atomic Force Microscopy. II. Hydrodynamic Force Measurements
Here we report a new study on the boundary conditions for the flow of a simple liquid in a confined geometry obtained by measuring hydrodynamic drainage forces with colloid probe atomic force microscopy (AFM). In this work, we provide experimental data obtained using a best practice experimental protocol and fitted with a new theoretical calculation (Zhu, L.; Attard, P.; Neto, C. Langmuir 2010, submitted for publication, preceding paper). We investigated the hydrodynamic forces acting on a silica colloid probe approaching a hydrophobized silicon surface in a single-component viscous Newtonian liquid (di-n-octylphthalate), a partially wetting system. The measured average slip lengths were in the range of 24-31 nm at approach velocities of between 10 and 80 mu m/s. Using our experimental approach, the presence of nanoparticle contaminants in the system can be indentified, which is important because it has been shown that nanoparticles lead to a large apparent slip length. Under our stringent control of experimental conditions, the measurement of the slip length is reproducible and independent of the spring constant of the cantilever.