Journal of Hazardous Materials, Vol.162, No.1, 44-49, 2009
Study on solid phase extraction and graphite furnace atomic absorption spectrometry for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent
A solid phase extraction and graphite furnace atomic absorption spectrometry (GFAAS) for the determination of nickel, silver, cobalt, copper, cadmium and lead with MCI GEL CHP 20Y as sorbent was studied. Trace amounts of chromium, nickel, silver, cobalt, copper, cadmium and lead were reacted with 2-(2-quinolinil-azo)-4-methyl-1,3-dihydroxidobenzene (QAMDHB) followed by adsorption onto MCI GEL CHP 20Y solid phase extraction column, and 1.0 mol L-1 HNO3 Was used as eluent. The metal ions in 300 mL solution can be concentrated to 1.0 mL, representing an enrichment factor of 300 was achieved. The recoveries of analytes at pH 8.0 with 1.0 g of resin were greater than 95% without interference from alkaline, earth alkaline and some metal ions. When detected with graphite furnace atomic absorption spectrometry, the detection limits in the original samples were 1.4 ng L-1 for Cr(III), 1.0 ng L-1 for Ni(II), 0.85 ng L-1 for Ag(I), 1.2 ng L-1 for Co(II), 1.0 ng L-1 for Cu(II), 1.2 ng L-1 for Cd(II) and 1.3 ng L-1 for Pb(II). The validation of the procedure was performed by the analysis of the certified standard reference materials, and the presented procedure was applied to the determination of analytes in biological, water and soil samples with good results (recoveries range from 89 to 104%, and R.S.D.% lower than 3.2%. The results agreed with the standard Value or reference method). (C) 2008 Published by Elsevier B.V.
Keywords:Solid phase extraction;Heavy metal ions;2-(2-Quinolinil-azo)-4-methyl-1,3-dihydroxidobenzene;Atomic absorption spectrometry;MCI GEL CHP 20Y resin