Electrophoresis, Vol.30, No.10, 1774-1782, 2009
Enantioselective determination of thyroxine enantiomers by ligand-exchange CE with UV absorbance and ICP-MS detection
A simple CE method has been developed for the separation and determination of thyroxine (T-4) enantiomers in pharmaceutical formulations. The method was based on ligand-exchange mechanism using a Cu(II)/L-proline complex as chiral selector. The effects of different parameters affecting separation such as chiral selector concentration, organic additive, buffer pH and temperature were investigated. A baseline separation of the two enantiomers was obtained at a Cu(II)/L-proline ratio of 1:8 in a borate buffer (15 mmol/L, pH 9.6) containing 10% v/v acetonitrile. Under the optimized conditions, precision linearity range and detection limits of the developed enantioselective CE method were evaluated and compared using two different detection systems: conventional UV detection at 226 nm and iodine (I-127) specific detection ("chiral speciation") with ICP-MS. Both methodologies show adequate analytical performance characteristics with detection limits around 0.30 mu g/mL for each enantiomer of T-4. Finally, a levothroid pharmaceutical formulation sample was successfully analyzed using both developed methods CE-UV and CE-ICP-MS.