화학공학소재연구정보센터
Electrophoresis, Vol.27, No.22, 4480-4486, 2006
Development of a NACE method for simultaneous measurement of three adenosine monophosphate isomers in biomimicking prebiotic synthesis without sample pretreatment
A practical NACE method was developed for simultaneous determination of three adenosine monophosphate (AMP) isomers. Separation of three AMP isomers was achieved using 200 mM Tris/H3BO3 in acetontrile/water (2:1 v/v) at pH* 10.0 as the running buffer and +25 kV as the applied voltage over a bare fused-silica capillary of 50 mu m id x 375 mu m od x 54.5 cm (46 cm to the detector window). At 260 nm, the calibration curves were linear in the range of 1-100 mu g/mL. The detection limits were less than 0.70 mu g/mL. The recovery ranged from 94.5 to 106.4%. The intraday RSDs of the migration times were between 2.1 and 3.0%. The developed NACE method has been successfully applied for the determination of three AMP isomers in the real samples of biomimicking prebiotic synthesis reaction between N-(O,O-diisopropyl) phosphoryl amino acid and adenosine.