Electrophoresis, Vol.24, No.12-13, 1935-1943, 2003
Flow injection-capillary electrophoresis system with contactless conductivity detection and hydrostatic pressure generated flow. Application to the quantitative analysis of inorganic anions in water samples
A simple and inexpensive flow injection-capillary electrophoresis (FI-CE) system with contactless conductivity detection (CCD) for automated quantitative analysis of chloride, nitrate, and sulfate in various water samples is demonstrated. A glass bottle containing the background electrolyte that is raised above the FI-CE interface generates a pulse-free, highly reproducible flow of the electrolyte through the FI-CE interface. The system operates at a flow rate of 300 muLmin(-1) with an injection volume of only 4 muL. The repeatability of peak areas (n = 18) was better than 0.81% RSD and the sample throughput was 90 samples per hour using the background electrolyte containing 12 MM L-histidine adjusted to pH 4.00 with acetic acid. The limits of detection were better than 125 mugL(-1) and were comparable to those obtained by conventional CE systems with CCD. Various calibration methods for FI-CE system with electrokinetic injection were tested and their suitability for the analysis of anions in real samples was evaluated.