화학공학소재연구정보센터
Electrophoresis, Vol.24, No.7-8, 1276-1280, 2003
Spotting and quantification of phosphoproteins purified by gel electrophoresis and laser ablation-element mass spectrometry with phosphorus-31 detection
Laser ablation inductively coupled plasma-mass spectrometry (ICP-MS) with P-31 detection has been used for spotting of phosphoproteins after one-dimensional polyacrylamide gel electrophoresis (1-D PAGE) and membrane transfer. By analyzing a mixture of myoglobin, alpha-casein and reduced fibrinogen it is demonstrated that phosphoproteins are specifically recognized by this method. A special washing step was found to be necessary to remove phosphate noncovalently bound to proteins. The P-31 signal was found to contain quantitative information both with respect to relative and absolute amounts of phosphorus present in phosphoproteins. Normalizing the P-31 signal from a single laser ablation trace by the total amount of phosphoprotein applied to the gel, a detection limit of 5 pmol of phosphorus is estimated.