화학공학소재연구정보센터
Electrophoresis, Vol.24, No.1-2, 185-192, 2003
Low-voltage electroosmosis pump for stand-alone microfluidics devices
Two types of low-voltage electroosmosis pumps were developed using microfabrication technology for usage in handy or stand-alone applications of the micrototal analysis systems (mu-TAS) and the lab-on-a-chip. This was done by making a thin (< 1 mum) region in the flow path and by only applying voltages near this thin region using electrodes inserted into the flow path. The inserted electrodes must be free from bubble formation and be gas-tight in order to avoid pressure leakage. For these electrodes, Ag/AgCl or a gel salt bridge was used. For patterning the gel on the chip, a hydrophilic photopolymerization gel and a photolithographic technique were optimized for producing a gel with higher electric conductivity and higher mechanical strength. For high flow rate application, wide (33.2 mm) and thin (400 nm) pumping channels were compacted into a 1 mm x 6 mm area by folding. This pump achieves an 800 Pa static pressure and a flow of 415 nL/min at 10 V. For high-pressure application, a pump was designed with the thin and thick regions in series and positive and negative electrodes were inserted between them alternatively. This pump could increase the pumping pressure without increasing the supply voltage. A pump with 10-stage connections generated a pressure of 25 kPa at 10 V.