화학공학소재연구정보센터
Biochemical and Biophysical Research Communications, Vol.326, No.1, 87-92, 2005
An enzymatic cycling method for the determination of serum total bile acids with recombinant 3 alpha-hydroxysteroid dehydrogenase
A highly sensitive enzymatic cycling method was developed for the serum total bile acids assay. We constructed a prokaryotic expression system to prepare the recombinant 3alpha-hydroxysteroid dehydrogenase in place of the natural enzyme and for the first time used it in the total bile acids assay. The production rate of thio-NADH correlated with the bile acids concentration and was measured by the change of absorbance at 405/660 nm. The enzymatic cycling method could detect 0.22 mumol/L total bile acids in serum. Within-run and between-run imprecisions were 1.2-3.7% and 2.3-4.8%, respectively. The calibration curve for total bile acids in serum was linear between 0.5 and 180 mumol/L. This method was free from interference by bilirubin, hemoglobin, ascorbate, and lactate dehydrogenase. In conclusion, serum total bile acids could be measured by the enzymatic cycling method with recombinant 3alpha-hydroxysteroid dehydrogenase as the tool enzyme. (C) 2004 Elsevier Inc. All rights reserved.