화학공학소재연구정보센터
Journal of Physical Chemistry B, Vol.110, No.6, 2949-2956, 2006
Specific anion effects on glass electrode pH measurements of buffer solutions: Bulk and surface phenomena
The effect of electrolytes on pH measurements via glass electrodes is explored with solutions buffered at pH 7 (phosphate and cacodylate). Salt and buffer concentrations are varied. Direct and reverse Hofmeister effects are observed. The phenomena are significant for salt concentrations above 0.1 M and for buffer concentrations below 20 mM. Changes in measured pH show LIP Most strongly with anions. They can be related to the usual physicochemical parameters (anion molar volumes, molar refractivity, and surface tensions) that are characteristic of Hofmeister series. They correlate strongly with anionic excess polarizabilities; this suggests the involvement of non-electrostatic, or dispersion, forces acting on ions. These forces contribute to ionic adsorption at the glass electrode surface, and to the liquid junction potential.