Turkish Journal of Chemistry, Vol.31, No.3, 293-299, 2007
Speciation of manganese using a pneumatic flow injection analysis-tandem spectrometer system
A pneumatic flow injection-tandem spectrometer system, without a delivery pump, was used for the speciation of manganese (Mn). In this system the suction force of the pneumatic nebulizer of a flame atomic absorption spectrometer (FAAS) was used for solution delivery through the manifold. Mn(VII) and total Mn (Mn(VII) and Mn(II)) concentrations were determined using a UV-Vis spectrometer and FAAS, respectively. The Mn(II) concentration was determined by the difference between the two. The calibration curves were linear tip to 15.00 mg L-1 for Mn(VII) and total Mn, with a detection limit of 0.08 mg L-1 and 0.05 mg L-1 for Mn(VII) and Mn, respectively. The mid-range precision and accuracy were < 1.89% and +/- 2.50% for the 2 species, respectively, at a sampling rate of 80 samplings h(-1). This system was used for the determination of Mn(VII) and Mn(II) in spiked and natural water, as well as industrial water.
Keywords:pneumatic flow injection analysis;tandem spectrometer;manganese;speciation;UV-Vis;spectrometer;flame atomic absorption spectrometer.