Thin Solid Films, Vol.565, 267-270, 2014
Mechanical properties of atomic force microscopy probes with deposited thin films
Deposition of thin films on atomic force microscopy (AFM) probes is a common technique that is used to either improve the light reflectance of the back side of the probe cantilevers or to modify the surface properties of the probe tips. However, this technique also affects the force constant and resonance frequency of the AFM cantilevers. The present work investigates theoretically this effect in the approximation of very thin films. The cantilever force constant changes due to the contribution of bending moment of the elastic force in the deposited film, while the resonant frequency changes due to the film contributions to the bending force and inertia of the cantilever. It is found that the relative variations of cantilever force constant and resonance frequency depend on the film to cantilever thickness, density and elasticity modulus ratios. This theoretical prediction is confirmed by the experimental investigations on mechanical properties of silicon cantilevers covered by diamond like carbon (DLC) and fluorocarbon thin films obtained by high power impulse magnetron sputtering depositions. Moreover, the effect of film depositions on the cantilever mechanical properties is used to determine the elasticity modulus and mass density of the deposited thin films. (C) 2014 Elsevier B.V. All rights reserved.
Keywords:Atomic force microscopy probes;Diamond like carbon thin film;Fluorocarbon thin film;High power impulse magnetron sputtering deposition;Film elasticity modulus;Film mass density