화학공학소재연구정보센터
Electrochimica Acta, Vol.82, 109-114, 2012
Modified working electrodes for electrochemical whole-cell microchips
The electrode geometry and material have a significant effect on the electrochemical biochip transduction of chemical signals into electrical current or voltage. In this work we focus on the working electrode aiming to improve the signal level of live cell sensors integrated on solid-state microchips. We present here a model and measurements describing the effect of the electrode material and dimensions on the biochip performance. The research hypothesis was that the electrode transduction efficiency increases as its effective area increases. Therefore, we investigated two methods to increase the electrode effective area: 3D structures and a polymer modified electrode. An electrochemical microchip was fabricated with a working electrode that was further modified, resulting in two structure types: 3D metallic pillar-based and polypyrrole-coated. The electrochemical performance of both modified electrodes was characterized and their utilization as working electrodes in whole-cell biochips for toxicity sensing was studied. Bio-detection efficiency analysis demonstrated higher biosensing performance for the metallic (e.g. Cu/Au) pillar-based microchip than for the polypyrrole-modified and the non-modified microchips. Therefore, we conclude that the enhanced signal of the modified geometry electrode is most probably due to the increased effective surface area and the improved charge transfer efficiency. (c) 2012 Elsevier Ltd. All rights reserved.